Proteus 235; Version: PTS-8 versions before PTS-8.7.2 — FDA record Z-0466-2022

Reported terminated

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Nov 29, 2021. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Proteus 235; Version: PTS-8 versions before PTS-8.7.2
Reason
Proton Therapy System (PTS) software can be started in the clinical environment (clinical user) while some processes are still running in the test environment (tcs user). Initiating the startup of the clinical processes in the environment is not prevented even if some processes of the test environment (tcs) are still running. It may lead to situations where the system is using test processes without notifying the clinical user. If the versions of the processes are different between both environments, this could have an impact on patient treatments.
Classification
Class II
Recalling firm
Ion Beam Applications S.A.
Quantity
5 devices
Agency-reported status
Terminated
Source record identifier
Z-0466-2022

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityPAT.108 (US), PAT.110 (US), PAT.113 (US), PAT.114 (EU), SAT.119 (US).

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

IN USA: VA, OK, NJ, TN OUS: Czech Republic.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Jan 19, 2022
Recall initiation date
Nov 29, 2021
Center classification date
Jan 7, 2022
Termination date
Feb 20, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records