DigitalDiagnost 4 (1) High Performance, (2) Flex/Value; Model Number: (1) 712031, (2) 712032; — FDA record Z-0464-2026

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Oct 1, 2025. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
DigitalDiagnost 4 (1) High Performance, (2) Flex/Value; Model Number: (1) 712031, (2) 712032;
Reason
Philips has identified that some ceiling suspension telescopic carriages manufactured from July 2024 through June 2025 may be difficult to move vertically (i.e. additional force/physical exertion required for manual movement) due to potential manufacturing issues.
Classification
Class II
Recalling firm
Philips Medical Systems DMC GmbH
Quantity
12 units
Agency-reported status
Ongoing
Source record identifier
Z-0464-2026

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicability1. Model Number: 712031; UDI-DI: 00884838065321; Serial Numbers: 10001181, 10001182, 10001185, 10001186, 10001188, 10001189, SN16000022, SN20000013; 2. Model Number: 712032; UDI-DI: 00884838074514; Serial Numbers: 10001179, 19000341, 19000342, 19000343;

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide Distribution: US (Nationwide) Distribution and OUS (International) countries of: Australia, Austria, Bahrain, Belgium, Bosnia-Herz., Canada, Chile, Croatia, Denmark, Estonia, Finland, France, Germany, Ghana, Gibraltar, Guadeloupe, Indonesia, Iraq, Israel, Italy, Jordan, Kuwait, Kyrgyzstan, Latvia, Lithuania, Netherlands, New Zealand, Norway, Panama, Philippines, Poland, Portugal, R¿union, Samoa, Saudi Arabia, Singapore, South Africa, South Korea, Spain, Sweden, Switzerland, T¿rkiye, Turkmenistan, United Kingdom, Utd.Arab.Emir.,

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Nov 19, 2025
Recall initiation date
Oct 1, 2025
Center classification date
Nov 7, 2025

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records