Convenience kits; DYNJ901047R ORTHO SPINE FUMICH DYNJ902308K TOTAL JOINT DYNJ909111F DR SALMA'S SPINE DYNJ909117G PAT... — FDA record Z-0451-2026

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Oct 10, 2025. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Convenience kits; DYNJ901047R ORTHO SPINE FUMICH DYNJ902308K TOTAL JOINT DYNJ909111F DR SALMA'S SPINE DYNJ909117G PATTERSON TOTAL JOINT DYNJ909113F KYPHOPLASTY
Reason
Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also escape into the operating room environment.
Classification
Class I
Recalling firm
Medline Industries, LP
Quantity
96,406 total units
Agency-reported status
Ongoing
Source record identifier
Z-0451-2026

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityDR SALMA'S SPINE DYNJ909111F UDI-DI 40198459130794 (case), 10198459130793 (ea) Lots 25FBE036 25GBV759; ORTHO SPINE FUMICH DYNJ901047R UDI-DI 40198459130466 (case), 10198459130465 (ea) Lots 25FBR230 25GBU693; PATTERSON TOTAL JOINT DYNJ909117G UDI-DI 40198459130480 (case), 10198459130489 (ea) Lots 25FBO307 25GBN625; TOTAL JOINT DYNJ902308K UDI-DI 40198459130596 (case), 10198459130595 (ea) Lots 25FBB321 25GBU688 KYPHOPLASTY DYNJ909113F UDI-DI 40198459130459 (case), 10198459130458 (ea) Lot 25IBM788

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

International distribution to Mexico and Panama. US nationwide distribution to the following: AK AL AR AZ CA CO CT FL GA IA IL IN KS KY MA ME MI MN MO MS MT NC NJ NY OH OR PA PR RI SC TN TX WA WI;

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Nov 19, 2025
Recall initiation date
Oct 10, 2025
Center classification date
Nov 13, 2025

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.