Recall archive / FDA device enforcement
Convenience kits; DYNJ901047R ORTHO SPINE FUMICH DYNJ902308K TOTAL JOINT DYNJ909111F DR SALMA'S SPINE DYNJ909117G PAT... — FDA record Z-0451-2026
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Oct 10, 2025. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- Convenience kits; DYNJ901047R ORTHO SPINE FUMICH DYNJ902308K TOTAL JOINT DYNJ909111F DR SALMA'S SPINE DYNJ909117G PATTERSON TOTAL JOINT DYNJ909113F KYPHOPLASTY
- Reason
- Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also escape into the operating room environment.
- Classification
- Class I
- Recalling firm
- Medline Industries, LP
- Quantity
- 96,406 total units
- Agency-reported status
- Ongoing
- Source record identifier
- Z-0451-2026
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | DR SALMA'S SPINE DYNJ909111F UDI-DI 40198459130794 (case), 10198459130793 (ea) Lots 25FBE036 25GBV759; ORTHO SPINE FUMICH DYNJ901047R UDI-DI 40198459130466 (case), 10198459130465 (ea) Lots 25FBR230 25GBU693; PATTERSON TOTAL JOINT DYNJ909117G UDI-DI 40198459130480 (case), 10198459130489 (ea) Lots 25FBO307 25GBN625; TOTAL JOINT DYNJ902308K UDI-DI 40198459130596 (case), 10198459130595 (ea) Lots 25FBB321 25GBU688 KYPHOPLASTY DYNJ909113F UDI-DI 40198459130459 (case), 10198459130458 (ea) Lot 25IBM788 |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
International distribution to Mexico and Panama. US nationwide distribution to the following: AK AL AR AZ CA CO CT FL GA IA IL IN KS KY MA ME MI MN MO MS MT NC NJ NY OH OR PA PR RI SC TN TX WA WI;
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
- Alaska
- Alabama
- Arkansas
- Arizona
- California
- Colorado
- Connecticut
- Delaware
- Florida
- Georgia
- Hawaii
- Iowa
- Idaho
- Illinois
- Indiana
- Kansas
- Kentucky
- Louisiana
- Massachusetts
- Maryland
- Maine
- Michigan
- Minnesota
- Missouri
- Mississippi
- Montana
- North Carolina
- North Dakota
- Nebraska
- New Hampshire
- New Jersey
- New Mexico
- Nevada
- New York
- Ohio
- Oklahoma
- Oregon
- Pennsylvania
- Rhode Island
- South Carolina
- South Dakota
- Tennessee
- Texas
- Utah
- Virginia
- Vermont
- Washington
- Wisconsin
- West Virginia
- Wyoming
Dates in the source
- Report date
- Nov 19, 2025
- Recall initiation date
- Oct 10, 2025
- Center classification date
- Nov 13, 2025
FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.