SYNCHRON Systems Direct LDL Cholesterol (LDLD) Reagent, REF: 969706 — FDA record Z-0441-2026

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Sep 17, 2025. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
SYNCHRON Systems Direct LDL Cholesterol (LDLD) Reagent, REF: 969706
Reason
Beckman Coulter has identified that there is an increased probability Synchron Systems Direct LDL Cholesterol (LDLD REF 969706) reagent lot M404166 may generate false high results. The issue was discovered internally and confirmed on August 19, 2025 by Beckman Coulter.
Classification
Class II
Recalling firm
Beckman Coulter Inc.
Quantity
2,932 units
Agency-reported status
Ongoing
Source record identifier
Z-0441-2026

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityREF: 969706. Lot: M404166. Expiration Date: 30Nov2025

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide - US Nationwide distribution in the states of AL, AR, CA, FL, GA, HI, IA, ID, KS, KY, LA, MA, MD, ME, MI, MO, MS, NC, NE, NJ, NM, NY, OH, OK, OR, PA, PR, QC, SC, TX, WA, WV and the country of Canada.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Nov 12, 2025
Recall initiation date
Sep 17, 2025
Center classification date
Nov 6, 2025

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.