BD Alaris infusion Pump Module 8100, REF: 8100; KIT, BEZEL ASSY 8100BD, REF: 49000437; KIT ASSY ONE PIECE BEZEL 8100,... — FDA record Z-0430-2026

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Oct 17, 2025. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
BD Alaris infusion Pump Module 8100, REF: 8100; KIT, BEZEL ASSY 8100BD, REF: 49000437; KIT ASSY ONE PIECE BEZEL 8100, REF: 49000270. Used with the following new and updated documents: BD Alaris Infusion System Inspection Safety Reminder, Part: BD-163124, Revision: 00; Cleaning and Disinfecting Procedure, Part: P00000864; Cleaning and Disinfecting Quick Reference Guide
Reason
If infusion pump is dropped or severely jarred this may damage the pump module bezel assembly, which can cause under-infusion, over-infusion, unregulated flow, or pump module failure to calibrate; so dropped/jarred pumps should be removed from use, tested, and inspected by qualified service personnel prior to reuse; tip sheet, cleaning and disinfecting procedure and reference guide updated.
Classification
Class I
Recalling firm
CareFusion 303, Inc.
Quantity
2,574,396
Agency-reported status
Ongoing
Source record identifier
Z-0430-2026

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityAll Alaris Pump Modules distributed after June 19, 2004. REF/UDI-DI: 8100/10885403810015, 10885403810046, 10885403810039, 10885403222054, 10885403517723, 10885403213212, 10885403211423; 49000437/10885403471018; 49000270/10885403247743

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide distribution. US nationwide including Puerto Rico, Guam, American Samoa, Northern Mariana Islands; and countries of SG, BH, AE, JP, CA, IT, QA, PH, DE, TW, PK, KR, BW, SA, MY, IL, NZ, AU, IN, GB, CN, NL, FR, ZA, BE, MX, CH, GI, KW, and TR.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Nov 26, 2025
Recall initiation date
Oct 17, 2025
Center classification date
Nov 20, 2025

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records