Tandem Mobi Insulin Pump: Description/Catalog: Tandem insulin pumps are battery-operated infusion pumps capable of bo... — FDA record Z-0427-2026
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Oct 6, 2025. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- Tandem Mobi Insulin Pump: Description/Catalog: Tandem insulin pumps are battery-operated infusion pumps capable of both basal and bolus delivery of insulin. The pumps utilize a motor-driven mechanism to deliver insulin from within a disposable cartridge, through an infusion set, into a patient s subcutaneous tissue. As with current insulin infusion pumps on the U.S. market, the desired timing and quantity of the insulin delivery is programmed by the user (i.e., the patient). Pump, Tandem Mobi, Control-IQ/1010750, Pump, Tandem Mobi, Control-IQ, Replacement/1012719, Pump, Tandem Mobi, Control-IQ 7.6.0.3, Medicare/1013501, Pump, Tandem Mobi, Control-IQ 7.7/1013655, Pump, Tandem Mobi, Control-IQ, 7.7, Replacement/1013656, Pump, Tandem Mobi, Control-IQ 7.7, Medicare/1013700, Pharmacy Kit, Starter Pack, Tandem Mobi 7.7/1014081
- Reason
- Insulin pump includes a vibration motor that gives tactile feedback for any alerts, alarms, or malfunctions. Pump may exhibit false vibration motor failure due to a software issue causing Malfunction 12: "Pump cannot operate, the mobile app can no longer receive data from the pump. Insulin delivery and any active CGM Sessions have been stopped", which could result in hyperglycemia.
- Classification
- Class I
- Recalling firm
- Tandem Diabetes Care, Inc.
- Quantity
- 17,745
- Agency-reported status
- Ongoing
- Source record identifier
- Z-0427-2026
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | Software versions: 7.6.0.1, 7.6.0.3, and 7.7.0.1 Catalog/UDI-DI: 1010750/00389152075013, 1012719/00389152271910, 1013501/00389152350110, 1013655/00389152365510, 1013656/00389152365619, 1013700/00389152370019, 1014081/00389152408170 |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
US: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, PT, RI, SC, SD, TN, TX, UT, VA, VI, VT, WA, WI, WV, WY
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
- Alaska
- Alabama
- Arkansas
- Arizona
- California
- Colorado
- Connecticut
- District of Columbia
- Delaware
- Florida
- Georgia
- Hawaii
- Iowa
- Idaho
- Illinois
- Indiana
- Kansas
- Kentucky
- Louisiana
- Massachusetts
- Maryland
- Maine
- Michigan
- Minnesota
- Missouri
- Mississippi
- Montana
- North Carolina
- North Dakota
- Nebraska
- New Hampshire
- New Jersey
- New Mexico
- Nevada
- New York
- Ohio
- Oklahoma
- Oregon
- Pennsylvania
- Rhode Island
- South Carolina
- South Dakota
- Tennessee
- Texas
- Utah
- Virginia
- Vermont
- Washington
- Wisconsin
- West Virginia
- Wyoming
Dates in the source
- Report date
- Nov 12, 2025
- Recall initiation date
- Oct 6, 2025
- Center classification date
- Nov 5, 2025
FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.