Omnipod 5 Automated Insulin Delivery System Product catalog numbers: PT-000409: ASM, Omnipod 5, PDM, Programmed PT-00... — FDA record Z-0423-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Nov 14, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Omnipod 5 Automated Insulin Delivery System Product catalog numbers: PT-000409: ASM, Omnipod 5, PDM, Programmed PT-000428: Adaptor, Charging, Noetic Nuu N5004L, Horizon PDM, US PT-000429: Cable, Charging, Omnipod 5, Horizon PDM
Reason
The firm has become aware of an issue with the Omnipod 5 Controller where the Controller charging port and charging cable are melting, deforming, or discoloring due to heat generated by increased resistance in the Controller port. The excess heat may cause minor burns if touched, and the Omnipod 5 Controller may fail to charge, requiring the user to switch to their backup insulin plan.
Classification
Class II
Recalling firm
Insulet Corporation
Quantity
129,086
Agency-reported status
Ongoing
Source record identifier
Z-0423-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityALL serial numbers/ controller lots are currently in scope of this recall

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Domestic distribution nationwide.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Dec 14, 2022
Recall initiation date
Nov 14, 2022
Center classification date
Dec 7, 2022

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.