Philips Allura Xper FD series with Software Version Number: 2.1.x- intended for Vascular, cardiovascular and neurovas... — FDA record Z-0416-2022

Reported terminated

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Nov 19, 2021. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Philips Allura Xper FD series with Software Version Number: 2.1.x- intended for Vascular, cardiovascular and neurovascular imaging applications, including diagnostic, interventional and minimally invasive procedures. Model Number: 722005, 722006, 722008, 722010, 722012, 722013, 722025, 722026, 722027, 722028.
Reason
Set screws securing the two gliding rods of the Extra Monitors (called 7th and 8th monitor) mounted on the Monitor Ceiling Suspension (MCS) can come loose, the Extra monitors might disengage from the Mounting Ceiling Suspension structure and fall down, may lead to patient, user or service engineer harm that may require medical intervention.
Classification
Class II
Recalling firm
Philips North America Llc
Quantity
USA: 5 systems R.O.W.: 15 systems
Agency-reported status
Terminated
Source record identifier
Z-0416-2022

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUS Serial Number/UDI: 1 (01)00884838054202(21)1; 527 (01)00884838059054(21)527; 408 (01)00884838054202(21)408; 306 (01)00884838054202(21)306; 61 (01)00884838059030(21)61 OUS: Model Product S/N UDI: 722006 Allura Xper FD20 1997 (01)00884838059191(10)1997; 722010 Allura Xper FD10 1177 (01)00884838059030(21)1177; 722010 Allura Xper FD10 1178 (01)00884838059030(21)1178; 722013 Allura Xper FD20 Biplane 544;; 722010 Allura Xper FD10 514 (01)00884838059030(21)514; 722013 Allura Xper FD20 Biplane 271; 722025 Allura Xper FD20 Biplane OR Table 6; 722005 Allura Xper FD10/10 519; 722008 Allura Xper FD20 Biplane 531; 722012 Allura Xper FD20 1901 (01)00884838059054(21)1901; 722012 Allura Xper FD20 2505 (01)00884838059054(21)2505; 722013 Allura Xper FD20 Biplane 137; 722026 Allura Xper FD10 439 (01)00884838054189(21)439; 722134 Field ext. Xper vascular systems R7.6 98736; 722027 Allura Xper FD10/10 172 (01)00884838054196(21)172;

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide distribution - US Nationwide distribution in the states of DC, GA, MA, TX, VA and the countries of Dominican Republic, INDIA, Japan, Poland, Spain, Taiwan.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Dec 29, 2021
Recall initiation date
Nov 19, 2021
Center classification date
Dec 22, 2021
Termination date
Feb 7, 2024