with Notch 22M/15F sold as part of Breas Medical MPV circuit 2 pc Kits: 006093, 006653, an accessory to ventilator de... — FDA record Z-0405-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Oct 13, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
with Notch 22M/15F sold as part of Breas Medical MPV circuit 2 pc Kits: 006093, 006653, an accessory to ventilator device, Vivo 45LS. intended for partial insertion into a patient's mouth (in between the lips) to facilitate access to the respiratory system.
Reason
Produced with unintended open slits on the side of the mouthpiece body, may result in insufficient ventilation in spontaneous breathing patients using MPV-VCV and MPV-PCV modes of ventilation
Classification
Class II
Recalling firm
Breas Medical, Inc.
Quantity
215 units
Agency-reported status
Ongoing
Source record identifier
Z-0405-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-DI: 05030267120505 NIV Angled Mouthpiece with Notch 22M/15F; 07321820060931 MPV Kit 006093; 07321820066537 MPV Kit 006653 MPV mouthpiece Lot Numbers: 32304444, 32305269 and 32306017

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Nationwide Foreign: Canada

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Dec 6, 2023
Recall initiation date
Oct 13, 2023
Center classification date
Nov 27, 2023

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.