What the record says

Product
NextAR Shoulder Surgical Technique, Shoulder NextAR RSA Platform, CT based computer-assisted surgical navigation platform used to perform a reverse shoulder arthroplasty on the glenoid
Reason
A warning has been added to the Surgical Technique for NextAR Shoulder, advising verification of suitable patient conditions prior to placement of the NextAR Shoulder Target Holder: WARNING Target fixation requires adequate coracoid bone stability. Osteoporosis, weakness of the coracoid process, or other deficiencies, could lead to adverse events including bone fracture. Verify that the coracoid bone and relevant patient conditions are suitable before proceeding.
Classification
Class II
Recalling firm
Medacta Usa Inc
Quantity
14 units
Agency-reported status
Terminated
Source record identifier
Z-0372-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityDocument Reference Code 99.81NEXTAR.12US (for US market only), Revision Numbers: Rev.02 and previous

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

AL, FL, NH, CO, IL, SC, AR

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Dec 14, 2022
Recall initiation date
Oct 27, 2022
Center classification date
Dec 2, 2022
Termination date
Oct 23, 2023

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.