FULL VISION INC. TRACKMASTER, TREADMILL GE T2100-ST1 110V, Model #317-07926GE — FDA record Z-0368-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Aug 31, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
FULL VISION INC. TRACKMASTER, TREADMILL GE T2100-ST1 110V, Model #317-07926GE
Reason
This has been identified a rare occurrence, under a specific scenario, where the circuit is found in a latch-up condition on the PCB that controls the safety function of the Magnetic Safety Tether, that when the magnet is pulled loose as during a fall, the treadmill may fail to stop.
Classification
Class II
Recalling firm
Full Vision Inc
Quantity
69 units
Agency-reported status
Ongoing
Source record identifier
Z-0368-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI/DI 00860176000668, Serial Numbers: GEDC-6632, GEDC-6633, GEDC-6634, GEDC-6635, GEDC-6636, GEDC-6637, GEDC-6638, GEDC-6639, GEDC-6640, GEDC-6641, GEDC-6642, GEDC-6643, GEDC-6644, GEDC-6645, GEDC-6646, GEDC-6647, GEDC-6648, GEDC-6649, GEDC-6650, GEDC-6651, GEDC-6652, GEDC-6653, GEDC-6654, GEDC-6655, GEDC-6656, GEDC-6657, GEDC-6658, GEDC-6659, GEDC-6660, GEDC-6661, GEDC-6662, GEDC-6663, GEDC-6664, GEDC-6665, GEDC-6666, GEDC-6667, GEDC-6668, GEDC-6669, GEDC-6670, GEDC-6671, GEDC-6672, GEDC-6673, GEDC-6674, GEDC-6675, GEDC-6676, GEDC-6677, GEDC-6678, GEDC-6679, GEDC-6680, GEDC-6681, GEDC-6755, GEDC-6756, GEDC-6757, GEDC-6758, GEDC-6759, GEDC-6760, GEDC-6761, GEDC-6762, GEDC-6763, GEDC-6764, GEDC-6839, GEDC-6840, GEDC-6841, GEDC-6842, GEDC-6843, GEDC-6844, GEDC-6845, GEDC-6846, GEDC-6847

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Wi, GA, NY, Mexico

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Dec 14, 2022
Recall initiation date
Aug 31, 2022
Center classification date
Dec 2, 2022