Assay: IMMULITE 2000 Intact PTH; Test Code: iPT; Siemens Material Number (SMN): (1) 10381441, (2) 10381442; Catalog N... — FDA record Z-0366-2026
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Sep 22, 2025. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- Assay: IMMULITE 2000 Intact PTH; Test Code: iPT; Siemens Material Number (SMN): (1) 10381441, (2) 10381442; Catalog Number: (1) L2KPP2, (2) L2KPP6;
- Reason
- The potential for falsely depressed Intact PTH patient results at the low end of the assay range, less than or equal to 50 pg/mL (less than or equal to 5.3 pmol/L), when using specific lots identified by the firm on the IMMULITE 2000/IMMULITE 2000 XPi systems. The bias is observed in both serum and plasma samples.
- Classification
- Class II
- Recalling firm
- Siemens Healthcare Diagnostics, Inc.
- Quantity
- 3318 units (169 US, 3149 OUS)
- Agency-reported status
- Ongoing
- Source record identifier
- Z-0366-2026
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | 1. Siemens Material Number (SMN): 10381441; Catalog Number: L2KPP2; UDI-DI: 00630414961910; Lot Number: 385, ***Expanded 2/18/2026*** 387, 388, 389; 2. Siemens Material Number (SMN): 10381442; Catalog Number: L2KPP6; UDI-DI: 00630414961927; Lot Number: 385, ***Expanded 2/18/2026*** 387, 388, 389; |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
Worldwide - US Nationwide distribution in the states of AL, CA, DC, FL, ID, IN, LA, MD, NC, NJ, NY, PA, SC and the countries of Argentina, Australia, Bolivia, Bosnia Herzeg, Brazil, Bulgaria, Canada, Chile, Colombia, Cyprus, Czech Republic, Ecuador, Egypt, France, Germany, Greece, Guatemala, Honduras, Hungary, India, Iran, Israel, Italy, Kosovo, Kuwait, Latvia, Lebanon, Macedonia, Mexico, Nepal, Netherlands, P.R. China, Pakistan, Palestinian Ter, Paraguay, Peru, Poland, Portugal, Qatar, Romania, Russian Fed., Serbia, Slovakia, Slovenia, South Africa, Spain, Taiwan, Tunisia, Turkey, Turkmenistan, U.A.E., Ukraine, Uruguay, Uzbekistan.
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
- Alaska
- Alabama
- Arkansas
- Arizona
- California
- Colorado
- Connecticut
- Delaware
- Florida
- Georgia
- Hawaii
- Iowa
- Idaho
- Illinois
- Indiana
- Kansas
- Kentucky
- Louisiana
- Massachusetts
- Maryland
- Maine
- Michigan
- Minnesota
- Missouri
- Mississippi
- Montana
- North Carolina
- North Dakota
- Nebraska
- New Hampshire
- New Jersey
- New Mexico
- Nevada
- New York
- Ohio
- Oklahoma
- Oregon
- Pennsylvania
- Rhode Island
- South Carolina
- South Dakota
- Tennessee
- Texas
- Utah
- Virginia
- Vermont
- Washington
- Wisconsin
- West Virginia
- Wyoming
Dates in the source
- Report date
- Nov 5, 2025
- Recall initiation date
- Sep 22, 2025
- Center classification date
- Oct 27, 2025
FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.