IMMULITE and IMMULITE 1000 Total Testosterone Assay- In vitro diagnostic use with the IMMULITE and IMMULITE 1000 Anal... — FDA record Z-0355-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Oct 25, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
IMMULITE and IMMULITE 1000 Total Testosterone Assay- In vitro diagnostic use with the IMMULITE and IMMULITE 1000 Analyzers for the quantitative measurement of total testosterone in serum and heparinized plasma.
Reason
Confirmed an average positive bias of 40% for testosterone concentrations of approximately 660 ng/dL (22.9 nmol/L) and above with the IMMULITE/IMMULITE 1000 Total Testosterone Assay when compared to the IMMULITE 2000/IMMULITE 2000 XPi Total Testosterone Assay. Two (2) lots of Total Testosterone assays for the IMMULITE/IMMULITE 1000, kit lots #515 and #516 were actively distributed to customers when the issue was discovered by Siemens, and they are subject to the correction or removal in this recall (867 total quantity of reagents distributed for each lot). The violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. In both males and female, testosterone results are not used in isolation but are interpreted in conjunction with clinical history and symptomology as well as with other laboratory testing (e.g. gonadotropins and other sex steroids). A delay in the diagnosis may be experienced and medical intervention, such as repeat testing may be required.
Classification
Class II
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Quantity
867 units
Agency-reported status
Ongoing
Source record identifier
Z-0355-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI: (01)00630414964782(10)0515(17)20221130; (01)00630414964782(10)0516(17)20221231 Lot Numbers: 515, Exp Date: 2022-11-30 516, Exp Date: 2022-12-31

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide Distribution: US (nationwide) to states of: AL, CA, FL, GA, IN, KY, NC, NE, OR, SD, TX, UT; and OUS (Foreign) to countries of: Canada, Mexico, Austria, Argentina, Austria, Bangladesh, Colombia, Brazil, Bulgaria, Costa¿Rica, Cyprus, Estonia, Germany, Greece, India, U.A.E., Italy, Kazakhstan, Latvia, P.R.¿China, Romania, Russian¿Fed., South Africa, Spain, United Kingdom, Uruguay and Singapore.¿¿¿¿

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Dec 14, 2022
Recall initiation date
Oct 25, 2022
Center classification date
Dec 2, 2022

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records