Trinity Biotech HbA1c (GHb) Contorls Kit, 500uL (Levels I & II), REF 01-04-0020, Part number 05-01-0040B, A1c analysis — FDA record Z-0350-2023
Reported terminated
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Oct 5, 2021. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- Trinity Biotech HbA1c (GHb) Contorls Kit, 500uL (Levels I & II), REF 01-04-0020, Part number 05-01-0040B, A1c analysis
- Reason
- The Control Level I material value was showing results exceeding the upper limit of the acceptable range after a period of three (3) months. Additionally, per the firm, the unacceptable Control Level 1 value results in the Analyzer generating an error code which stops further testing and nullifies associated patient samples, resulting in a delay in patient results and a potential delay in therapeutic modifications.
- Classification
- Class III
- Recalling firm
- Primus Corporation dba Trinity Biotech
- Quantity
- 182 kits
- Agency-reported status
- Terminated
- Source record identifier
- Z-0350-2023
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | UDI/DI 0539151674808, Lot 12161 |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
Turkey, India, Romania, Thailand, Vietnam, S. Korea, Peru, Pakistan Austria, UAE, Chile, Italy.
No US distribution state could be established from this record. Check the official notice.
Dates in the source
- Report date
- Dec 7, 2022
- Recall initiation date
- Oct 5, 2021
- Center classification date
- Dec 1, 2022
- Termination date
- Aug 9, 2023