Puritan Bennett 980 Series Ventilator, 980X1ENDIUU — FDA record Z-0347-2022

Reported terminated

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Nov 4, 2021. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Puritan Bennett 980 Series Ventilator, 980X1ENDIUU
Reason
Manufacturing assembly error where a capacitor within the ventilator was assembled incorrectly, which may cause the device to become inoperable during use.
Classification
Class I
Recalling firm
Covidien Llc
Quantity
278 total
Agency-reported status
Terminated
Source record identifier
Z-0347-2022

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityGTIN/UDI - 10884521171541 Serial Numbers: 35B1700300 35B1700306 35B1700307 35B1700362 35B1700372 35B1700384 35B1700387 35B1700388 35B1700396 35B1700397 35B1700403 35B1700413 35B1700416 35B1700418 35B1700426 35B1700431 35B1700434 35B1700440 35B1700445 35B1700451 35B1700452 35B1700453 35B1700459 35B1700461 35B1700464 35B1700468 35B1700472 35B1700477 35B1700478 35B1700479 35B1700480 35B1700484 35B1700485 35B1700494 35B1700495 35B1700496 35B1700501 35B1700502 35B1700503 35B1700504 35B1700534 35B1700574 35B1700579 35B1700604 35B1700609 35B1700638 35B1700655 35B1700664 35B1700816 35B1700919 35B1700994 35B1701001 35B1701418 35B1701698 35B1702832 35B1703438 35B1800033 35B1800721 35B1800811

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide distribution - US Nationwide.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Dec 15, 2021
Recall initiation date
Nov 4, 2021
Center classification date
Dec 9, 2021
Termination date
Jul 9, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.