Accelerate PhenoTest BC kit REF 10101018 — FDA record Z-0345-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Oct 17, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Accelerate PhenoTest BC kit REF 10101018
Reason
Due improperly formulated and released bulk lot of fluorescence in situ hybridization (FISH) probes resulting in false positive identification results for Staphylococcus aureus and Streptococcus spp.
Classification
Class II
Recalling firm
Accelerate Diagnostics Inc
Quantity
200 kits
Agency-reported status
Ongoing
Source record identifier
Z-0345-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityPart Number: 10101018 UDI-DI Code: 00862011000307 Lot Numbers: 5549A 5566A 5588A 5606A 5612A

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

U.S.: AR, CA, DC, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MN, MO, MS, MT, NC, ND, NJ, NY, OH, PA, RI, SC, TN, TX, VA, WA, and WV. O.U.S.: Germany, Italy, Kuwait, Latvia, Poland, Portugal, Romania, Saudi Arabia, Spain, and United Arab Emirates

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Dec 7, 2022
Recall initiation date
Oct 17, 2022
Center classification date
Nov 30, 2022