Cardiosave Hybrid IntraAortic Balloon Pump. Model Numbers: 0998-00-0800-31, 0998-UC-0800-31, 0998-00-0800-32, 0998-UC... — FDA record Z-0336-2022
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Nov 15, 2021. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- Cardiosave Hybrid IntraAortic Balloon Pump. Model Numbers: 0998-00-0800-31, 0998-UC-0800-31, 0998-00-0800-32, 0998-UC-0800-32, 0998-00-0800-33, 0998-UC-0800-33, 0998-00-0800-34, 0998-UC-0800-34, 0998-00-0800-35, 0998-UC-0800-35, 0998-00-0800-36, 0998-UC-0800-36, 0998-00-0800-45, 0998-UC-0800-45, 0998-00-0800-52, 0998-UC-0800-52, 0998-00-0800-53, 0998-UC-0800-53, 0998-00-0800-55, 0998-UC-0800-55, 0998-00-0800-65, 0998-UC-0800-65.
- Reason
- Firm is initiating a correction due to the possibility of fluid ingress. Fluid entering the device may short various electronic components thus leading to system shutdown, which could result in unsafe hemodynamic instability.
- Classification
- Class I
- Recalling firm
- Datascope Corp.
- Quantity
- 5,566 (US); 6,354 (OUS)
- Agency-reported status
- Ongoing
- Source record identifier
- Z-0336-2022
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | All Serial Numbers. May also be designated as part number D998 etc. Model Number, UDI-DI: 0998-00-0800-31, 10607567109053; 0998-UC-0800-31, 10607567109053; 0998-00-0800-32, 10607567111117; 0998-UC-0800-32, 10607567111117; 0998-00-0800-33, 10607567109008; 0998-UC-0800-33, 10607567109008; 0998-00-0800-34, 10607567111940; 0998-UC-0800-34, 10607567111940; 0998-00-0800-35, 10607567109107; 0998-UC-0800-35, 10607567109107; 0998-00-0800-36, 10607567114187; 0998-UC-0800-36, 10607567114187; 0998-00-0800-45, 10607567108421; 0998-UC-0800-45, 10607567108421; 0998-00-0800-52, 10607567108438; 0998-UC-0800-52, 10607567108438; 0998-00-0800-53, 10607567108391; 0998-UC-0800-53, 10607567108391; 0998-00-0800-55, 10607567108414; 0998-UC-0800-55, 10607567108414; 0998-00-0800-65, 10607567113432; 0998-UC-0800-65, 10607567113432 |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
Worldwide distribution - US Nationwide.
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
- Alaska
- Alabama
- Arkansas
- Arizona
- California
- Colorado
- Connecticut
- Delaware
- Florida
- Georgia
- Hawaii
- Iowa
- Idaho
- Illinois
- Indiana
- Kansas
- Kentucky
- Louisiana
- Massachusetts
- Maryland
- Maine
- Michigan
- Minnesota
- Missouri
- Mississippi
- Montana
- North Carolina
- North Dakota
- Nebraska
- New Hampshire
- New Jersey
- New Mexico
- Nevada
- New York
- Ohio
- Oklahoma
- Oregon
- Pennsylvania
- Rhode Island
- South Carolina
- South Dakota
- Tennessee
- Texas
- Utah
- Virginia
- Vermont
- Washington
- Wisconsin
- West Virginia
- Wyoming
Dates in the source
- Report date
- Dec 22, 2021
- Recall initiation date
- Nov 15, 2021
- Center classification date
- Dec 15, 2021
FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.