Thermo SCIENTIFIC, Sensititre CMC7AFLF, Gram Negative IVD AST — FDA record Z-0332-2023

Reported terminated

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Oct 20, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Thermo SCIENTIFIC, Sensititre CMC7AFLF, Gram Negative IVD AST
Reason
An internal technical investigation has confirmed that Thermo Scientific" Gram Negative IVD AST Sensititre Plate may lead to false susceptible results for Proteae Tribe (including Proteus spp., Providencia spp., Morganella Morganii) when tested with doripenem, ertapenem, meropenem, imipenem, cefepime, piperacillin/tazobactam and aztreonam.
Classification
Class I
Recalling firm
Remel Inc
Quantity
1647 units
Agency-reported status
Terminated
Source record identifier
Z-0332-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityLot Numbers: B2281, B2151A, B2034, B1453

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide Distribution

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Dec 14, 2022
Recall initiation date
Oct 20, 2022
Center classification date
Dec 7, 2022
Termination date
Sep 30, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.