Medtronic Mo.Ma Ultra Proximal Cerebral Protection Device, Model Number MUS0130069X6 — FDA record Z-0320-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Oct 12, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Medtronic Mo.Ma Ultra Proximal Cerebral Protection Device, Model Number MUS0130069X6
Reason
The Mo.Ma Ultra device manifold is labelled with 2 stickers, an inferior and superior sticker, one on each side of the manifold. These stickers label the common carotid artery (CCA) proximal balloon and external carotid artery (ECA) distal balloon inflation and deflation ports. Both sides of the complaint units manifold were labelled with the superior label leading to an incorrect identification of the proximal and distal inflation /deflation ports when viewing the devices from the inferior side.
Classification
Class II
Recalling firm
Medtronic Inc
Quantity
2767 units
Agency-reported status
Ongoing
Source record identifier
Z-0320-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI/DI 08033477090719: UDI/DI 00763000451691

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide distribution - US Nationwide and the countries of Algeria, Armenia, France, Germany, Iraq, Italy, Japan, Netherlands, Poland, Romania, Spain, Sweden, Switzerland, United Kingdom.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Nov 29, 2023
Recall initiation date
Oct 12, 2023
Center classification date
Nov 17, 2023