The Access Cortisol assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination... — FDA record Z-0313-2025

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Sep 30, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
The Access Cortisol assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of cortisol levels in human serum, plasma (heparin, EDTA) and urine using the Access Immunoassay Systems.
Reason
The Access Cortisol assay protocol file (APF) applies an incorrect calibration curve acceptance parameter at the S1 calibrator level that can lead to a high rate of calibration failures which can vary by reagent lot number. Calibration failure would lead to delay of measurement results reporting and potential injuries.
Classification
Class II
Recalling firm
Beckman Coulter, Inc.
Quantity
149 units
Agency-reported status
Ongoing
Source record identifier
Z-0313-2025

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI/DI 15099590231118, all lots

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide distribution.

No US distribution state could be established from this record. Check the official notice.

Dates in the source

Report date
Nov 13, 2024
Recall initiation date
Sep 30, 2024
Center classification date
Nov 6, 2024