The Access Cortisol assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination... — FDA record Z-0313-2025
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Sep 30, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- The Access Cortisol assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of cortisol levels in human serum, plasma (heparin, EDTA) and urine using the Access Immunoassay Systems.
- Reason
- The Access Cortisol assay protocol file (APF) applies an incorrect calibration curve acceptance parameter at the S1 calibrator level that can lead to a high rate of calibration failures which can vary by reagent lot number. Calibration failure would lead to delay of measurement results reporting and potential injuries.
- Classification
- Class II
- Recalling firm
- Beckman Coulter, Inc.
- Quantity
- 149 units
- Agency-reported status
- Ongoing
- Source record identifier
- Z-0313-2025
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | UDI/DI 15099590231118, all lots |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
Worldwide distribution.
No US distribution state could be established from this record. Check the official notice.
Dates in the source
- Report date
- Nov 13, 2024
- Recall initiation date
- Sep 30, 2024
- Center classification date
- Nov 6, 2024
FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.