SHIP KIT, PROCEDURE KIT, PROTEKDUO LS - DL31 for temporary Cardiopulmonary bypass. Including the following components... — FDA record Z-0311-2022

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Oct 14, 2021. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
SHIP KIT, PROCEDURE KIT, PROTEKDUO LS - DL31 for temporary Cardiopulmonary bypass. Including the following components: a) 31 Fr ProtekDuo Veno-Venous Cannula Set b) LifeSPARC Pump c) 14Fr to 26 Fr Dilator Set d) TandemHeart Wrap. Kit Number: 5820-3118 Intended use is applicable to the intended use of specific devices included in the kit.
Reason
Top label of the outer package of the collection kit Item 5820-3118 showed expiration date of 2022/05/01 (correct expiration date) while side label showed expiration date of 2021/05/01 (incorrect expiration date)
Classification
Class III
Recalling firm
Cardiac Assist, Inc
Quantity
7 units
Agency-reported status
Ongoing
Source record identifier
Z-0311-2022

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilitySerial Numbers: 872170, 872171, 872172, 872173, 872182, 872183, 872184. UDI: 01)00814112020791(17)220 501(22)220511(21) 872170; (01)00814112020791(17)220 501(22)220511(21) 872171; (01)00814112020791(17)220 501(22)220511(21) 872172; (01)00814112020791(17)220 501(22)220511(21) 872173; (01)00814112020791(17)220 501(22)220511(21) 872182; (01)00814112020791(17)220 501(22)220511(21) 872183; 01)00814112020791(17)220 501(22)220511(21) 872184

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Distribution to states of: DC, MA, SC, and VA.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Dec 8, 2021
Recall initiation date
Oct 14, 2021
Center classification date
Nov 29, 2021

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.