AU/DxC AU Chemistry Transferrin REF: OSR6152 Immune complexes formed in solution scatter light in proportion to their... — FDA record Z-0309-2024
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Sep 6, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- AU/DxC AU Chemistry Transferrin REF: OSR6152 Immune complexes formed in solution scatter light in proportion to their size, shape and concentration. Turbidimeters measure the reduction of incident light due to reflection, absorption, or scatter. In the procedure, the measurement of the decrease in light transmitted (increase in absorbance) through particles suspended in solution as a result of complexes formed during the antigen-antibody reaction, is the basis of this assay.
- Reason
- The firm has identified that the Transferrin reagent does not meet the Instructions for Use (IFU) reagent on-board stability claims, which may cause control failures and/or erroneous patient samples results. The Complaint handling unit has received a total of 2 safety cases (CASE-2023-02027212 and CASE-2023-02045278) and 99 quality cases with the failure mode attributed to Transferrin reagent (OSR6152 / lot 2573) associated with reagent blank shifts for the Reagent 2 (R2), imprecision, calibration failures, erratic Quality Control (QC) recoveries, and/or erroneous high/low (serum, plasma, urine) patient results. The issue was confirmed through internal testing by Beckman Coulter on June 30, 2023. Reagent Blank shifts for R2, imprecision, calibration failures, erratic Quality Control (QC) recoveries, and/or erroneous high/low (serum, plasma, urine) patient results. The issue was confirmed through internal testing by Beckman Coulter on June 30, 2023.
- Classification
- Class II
- Recalling firm
- Beckman Coulter Biomedical Division
- Quantity
- 2,754
- Agency-reported status
- Ongoing
- Source record identifier
- Z-0309-2024
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | UDI-DI: 15099590010492, LOT: 2573, Expiration: 01 August 2024 |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
Worldwide - US Nationwide distribution including in the states of NY, CA, MS, GA, PA, SC, WA, TN, AL, AK, OR, MO, KS, FL, LA, OH, TX, NJ, MI, WV, MD, IL, RI, VA, NC, IN, OK, NV, NE, WY, KY, WI, ND, MA, NH, UT, CO, GU, PR, AR, AZ, ME and the countries of Canada, Netherlands, United Kingdom, Sweden, Germany, France, Switzerland, Tunisia, Australia, Belgium, Costa Rica, Uruguay, Turkey, South Africa, New Zealand, Italy, Thailand, Ghana, Russia, Hungary, Czech Republic, Slovakia, India, Austria, Poland, Romania, Ireland, Spain, Portugal, Mexico, Vietnam, Korea, Mozambique, Kazakhstan, China, Malaysia, Singapore, Kenya, Brazil, Namibia, Egypt, Panama.
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
- Alaska
- Alabama
- Arkansas
- Arizona
- California
- Colorado
- Connecticut
- Delaware
- Florida
- Georgia
- Hawaii
- Iowa
- Idaho
- Illinois
- Indiana
- Kansas
- Kentucky
- Louisiana
- Massachusetts
- Maryland
- Maine
- Michigan
- Minnesota
- Missouri
- Mississippi
- Montana
- North Carolina
- North Dakota
- Nebraska
- New Hampshire
- New Jersey
- New Mexico
- Nevada
- New York
- Ohio
- Oklahoma
- Oregon
- Pennsylvania
- Rhode Island
- South Carolina
- South Dakota
- Tennessee
- Texas
- Utah
- Virginia
- Vermont
- Washington
- Wisconsin
- West Virginia
- Wyoming
Dates in the source
- Report date
- Nov 22, 2023
- Recall initiation date
- Sep 6, 2023
- Center classification date
- Nov 16, 2023
FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.