DeRoyal CHILDREN'S GENERAL TRACEPACK, REF 89-9342-06 — FDA record Z-0307-2025

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Oct 3, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
DeRoyal CHILDREN'S GENERAL TRACEPACK, REF 89-9342-06
Reason
DeRoyal is recalling certain lots of surgical procedure pack products that contain a Stryker SafeAir Smoke Evacuation Pencil. The recall is due to Stryker receiving reports that the pencil is activating without manual input when plugged into a power source or remains active after buttons are released.
Classification
Class II
Recalling firm
DeRoyal Industries Inc
Quantity
160 units
Agency-reported status
Ongoing
Source record identifier
Z-0307-2025

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI/ DI 00749756916646 (each), 50749756916641 (case): Lot Numbers: lot 59839716 exp 12/1/2025

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution in the state of OH.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Nov 13, 2024
Recall initiation date
Oct 3, 2024
Center classification date
Nov 5, 2024