The Ocean Water Seal Chest Drain is a plastic, three bottle vacuum collection device. It consists of a collection cha... — FDA record Z-0307-2022

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Oct 22, 2021. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
The Ocean Water Seal Chest Drain is a plastic, three bottle vacuum collection device. It consists of a collection chamber that holds shed mediastinal blood, a water seal chamber using a water seal that prevents air from entering into the pleural space and a suction control chamber, which incorporates a mechanical regulator to control suction pressures imposed upon the patient's pleural cavity.
Reason
Inadequate/inappropriate Set-Up Instructions, which may result in delay in treatment and an unreasonable risk of harm to the patient
Classification
Class II
Recalling firm
Atrium Medical Corporation
Quantity
87,972 drains (14,662 cases; 6 units/case)
Agency-reported status
Ongoing
Source record identifier
Z-0307-2022

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityProduct Code/REF Number 2002-400; Lot codes: 441533, 443300, 443608, 443629, 444473, 445527, 448966, 449683, 450899, 450900, 451270, 451988, 452177, 452814, 454577, 455871, 458264, 458604, 459198, 459642, 462192, 462194, 464075, 464076, 465037, 466247, 469921, 470627, 471651,471703, 472562, 472785 and 473725. UDI Code: 00650862100211

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide Distribution Foreign: Bahrain Belgium Brunei Darussalam Canada Colombia Ecuador Finland Italy Kuwait Libya Mauritius Mexico Netherlands Poland Portugal Russian Federation Saudi Arabia Spain United Arab Emirates United Kingdom of Great Britain and Northern Ireland

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Dec 1, 2021
Recall initiation date
Oct 22, 2021
Center classification date
Nov 25, 2021

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records