Artis Q.zen floor with software VD12-angiography systems developed for single and biplane diagnostic imaging and inte... — FDA record Z-0303-2022

Reported terminated

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Oct 20, 2021. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Artis Q.zen floor with software VD12-angiography systems developed for single and biplane diagnostic imaging and interventional procedure Model Number: 10848353
Reason
If SID (source-to-image distance) lift movement is activated and x-ray shall be released simultaneously; x-ray is not possible and the message No x-ray, try again is displayed, may result in a short delay in procedure
Classification
Class II
Recalling firm
Siemens Medical Solutions USA, Inc
Quantity
3 units
Agency-reported status
Terminated
Source record identifier
Z-0303-2022

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilitySerial Numbers: 105500 105503 105501

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide Distribution

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Dec 1, 2021
Recall initiation date
Oct 20, 2021
Center classification date
Nov 24, 2021
Termination date
Jun 18, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.