IBA Proton Therapy System - PROTEUS 235. PTS-11 versions, PTS-12 versions with Orion PPS version OSS6.0. Configured w... — FDA record Z-0295-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Sep 18, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
IBA Proton Therapy System - PROTEUS 235. PTS-11 versions, PTS-12 versions with Orion PPS version OSS6.0. Configured with Orion Patient Positioning System.
Reason
The patient positioning system can perform motions in unexpected directions, creating the potential for crushing of the patient between the patient positioning devices.
Classification
Class II
Recalling firm
Ion Beam Applications S.A.
Quantity
5 (US), 9 (OUS)
Agency-reported status
Ongoing
Source record identifier
Z-0295-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-DI: (01)05404013801138. Serial numbers (US): SAT133, SAT140, SBF113, SBF128, SBF135 & (OUS): SAT125, SAT127, SAT132, SAT136, SBF107, SBF112, SBF117, SBF124, & SBF125.

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide - US Nationwide distribution in the states of Arkansas, Florida, Kansas, and Virginia and the countries of Belgium, France, India, Italy, Netherlands, Spain, Singapore, and Taiwan.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Nov 22, 2023
Recall initiation date
Sep 18, 2023
Center classification date
Nov 14, 2023