Atrium Advanta V12 Covered Stent System, 10mm x 59mm x 80cm (OUS)-Indicated for restoring and improving the patency o... — FDA record Z-0291-2024

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Sep 21, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Atrium Advanta V12 Covered Stent System, 10mm x 59mm x 80cm (OUS)-Indicated for restoring and improving the patency of the iliac andrenal arteries. Renal approval includes 5mm, 6mm and 7mm diameter Advanta V12 covered stent sizes. In Canada, the Advanta V12 Covered Stent indication excludes renal arteries. REF: 85361
Reason
Balloon-to-catheter weld failure that may cause loss of balloon pressure during stent deployment.
Classification
Class II
Recalling firm
Atrium Medical Corporation
Quantity
141 units
Agency-reported status
Ongoing
Source record identifier
Z-0291-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-DI: 00650862853612 Lot Numbers: 464909, 464910

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Foreign: Australia, Belgium, Canada, France, Germany, Greece, Malta, Netherlands, New Zealand, Portugal, and United Kingdom

No US distribution state could be established from this record. Check the official notice.

Dates in the source

Report date
Nov 22, 2023
Recall initiation date
Sep 21, 2023
Center classification date
Nov 10, 2023