epoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System, SMN 10736515 — FDA record Z-0286-2025

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Sep 24, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
epoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System, SMN 10736515
Reason
Siemens Healthcare Diagnostics Inc. has confirmed that a negative sodium bias may exist in one lot of epoc BGEM BUN Test Cards approximately 2 months prior to expiration. The observed maximum negative bias from internal testing is -10 mmol/L. No positive bias has been observed. Quality Control (QC) fluid may or may not detect the issue depending on the bias magnitude and the frequency of QC testing. One customer complaint has been associated with this issue. No adverse events have been reported. All affected lots have recently expired before the issuance of the recall.
Classification
Class II
Recalling firm
Siemens Healthcare Diagnostics Inc
Quantity
1292 units
Agency-reported status
Ongoing
Source record identifier
Z-0286-2025

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-DI: 00809708121860; Lot Numbers: 08-24092-60 08-24093-60 08-24094-60 08-24095-60 08-24096-60

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide. Canada.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Nov 13, 2024
Recall initiation date
Sep 24, 2024
Center classification date
Nov 1, 2024