Recall archive / FDA device enforcement
Brand Name: Vial2Bag Advanced¿ Product Name: Vial2Bag Advanced¿ 20mm Admixture Device Model/Catalog Number: 6070030 S... — FDA record Z-0283-2025
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Sep 24, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- Brand Name: Vial2Bag Advanced¿ Product Name: Vial2Bag Advanced¿ 20mm Admixture Device Model/Catalog Number: 6070030 Software Version: Not applicable Product Description: The Vial2Bag Advanced 20mm Admixture Device is a single use, fluid transfer device that allows for the reconstitution and transfer of fluids from drug vials into the IV bag containing infusion solution, through the IV bag administration port. The device consists of the body, Protector, IV Port, and an integrated vial adapter. The device is provided as a sterile, non-pyrogenic product. The device is intended to be used with standard drug vials with a seal diameter of 20mm and an elastomeric stopper. The Vial2Bag Advanced 20mm Admixture Device is designed to work with a standard 50, 100, or 250mL IV bag and an external IV administration set. The device does not contain any medicinal substances and there are no additional accessories needed or provided with the Vial2Bag Advanced 20mm Admixture Device for the device to meet its intended purpose. Component: No
- Reason
- Due to packaging damage from heat exposure during shipping/transportation that may result in a breach in the sterile barrier.
- Classification
- Class II
- Recalling firm
- West Pharmaceutical Services AZ, Inc.
- Quantity
- 129600
- Agency-reported status
- Ongoing
- Source record identifier
- Z-0283-2025
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | Lot Code: Recall Product 1: Product Code 6080030, Lot# K188, UDI (01)10850022888011(17)270430(10)K188\F(90)36098150\F(91)6070030 Recall Product 2: Product Code 6080030, Lot# K194, UDI (01)10850022888011(17)270430(10)K194\F(90)36098150\F(91)6070030 |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
A U.S. Distributor, Progressive Medical Inc. (PMI) currently supplies the Vial2Bag Advanced 20mm Admixture Device to hospitals in the U.S. (end user). PMI distributed the impacted Vial2Bag Advanced 20mm Admixture Device lots to the end users. U.S.: Distributor located in MO U.S. Customers: AL, AR, AZ, CA, CO, CT, FL, GA, HI, IL, IN, KY, LA, MD, MI, MN, MO, MS, NC, NE, NM, NV, NY, OH, OK, OR, PA,SC, TN, TX, UT, WA, and WI.
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
Dates in the source
- Report date
- Nov 6, 2024
- Recall initiation date
- Sep 24, 2024
- Center classification date
- Oct 31, 2024
FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.