Recall archive / FDA device enforcement
Alinity m System, REF 08N53-002, For In Vitro Diagnostic Use — FDA record Z-0268-2023
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Oct 12, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- Alinity m System, REF 08N53-002, For In Vitro Diagnostic Use
- Reason
- Abbott has identified potential performance issues for the Alinity m System software version 1.6.5: 1) In a unique scenario, the waste chute flapper was found to not open when the Systems Solution drawer was closed and locked. It was found that the software will cause the flapper door to stay in the closed position when the drawer is closed and locked if the flapper had previously been moved out of position. 2) Sample preparation drawer #1 barcode information is used instead of sample prep drawer #2 when the scanned data is not sent to the System Control Center before the next bottle barcode in sample prep drawer#2 is scanned. 3) Under a specific condition, while the system is processing tests and a new test request is made, when the level of the bulk solution is too low to process a test, the software would error stop the system and try to complete all in-process tests. An internal counter is reset which can cause reaction vessels (RVs) already present in an Amplification Detection Unit (or amp detect) to remain there and not be moved to waste so that more RVs can be placed into the amp detect for processing. 4) 4 to 6 replicates of the same auto calibrator orders can be run with 2 different sets of calibrator materials on the same rack. It was discovered during internal testing, when running 4 to 6 replicates of a calibrator, it is possible for the user to use 2 separate lots of material which is not detected by the System Control Center (SCC). If this were to occur, the calibration curve would be created using 2 separate material lots. In normal use scenario, this curve would be made using only 1 lot of material.
- Classification
- Class II
- Recalling firm
- Abbott Molecular, Inc.
- Quantity
- 883 units
- Agency-reported status
- Ongoing
- Source record identifier
- Z-0268-2023
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | All serial numbers |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
Worldwide distribution.
No US distribution state could be established from this record. Check the official notice.
Dates in the source
- Report date
- Nov 30, 2022
- Recall initiation date
- Oct 12, 2022
- Center classification date
- Nov 18, 2022