1) Flex Extenders, 5.2" x 15mm OD x 15mm ID REF S6411P 2) Flex Extender, 5.2" x 15mm OD x 22/15mm ID REF 7 5132 The W... — FDA record Z-0263-2025

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Sep 27, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
1) Flex Extenders, 5.2" x 15mm OD x 15mm ID REF S6411P 2) Flex Extender, 5.2" x 15mm OD x 22/15mm ID REF 7 5132 The Westmed Flex Extenders are non invasive, single use/disposable, large bore, corrugated plastic tubes that provide a flexible, low resistance lightweight connection between breathing circuits and patient interface (tracheal tube, tracheostomy tube, or mask).
Reason
Due to connector within the circuit flex extender not meeting specifications resulting in a low connection force and as a result, the connector may leak or disconnect and may cause life threating conditions or death.
Classification
Class II
Recalling firm
Westmed, Inc.
Quantity
50,050 units
Agency-reported status
Ongoing
Source record identifier
Z-0263-2025

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityCatalog Number: S6411P UDI-DI codes: Each: 00709078012488 Case: 10709078012485 Lot Numbers: 0004294660 0004289136 0004291318 0004295361 Catalog Number: 7 5132 UDI-DI codes: Each: 00709078008405 Case: 10709078008402 Lot Numbers: 0004289133 0004289134 0004289135 0004289735 0004289736 0004289737 0004290327 0004290328 0004291315 0004291316 0004291317 0004291790 0004291791 0004292272 0004292273 0004292275 0004292717 0004293127 0004294432 0004294433 0004295069 0004295358 0004295458

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

U.S.: AK, CA, CO, DC, FL, GA, IL, IN, KY, MA, MD, ME, MI, MN, MO, NC, NE, NH, NJ, NV, NY, OH, OR, PA, SC, TX, VA, VT, WA, and WV O.U.S.: N/A

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Nov 6, 2024
Recall initiation date
Sep 27, 2024
Center classification date
Oct 30, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.