Abbott Alinity m HPV AMP Kit, used with the Alinity m System, product codes: a) REF 09N15-095; b) REF 09N15-090; c) R... — FDA record Z-0233-2025

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Sep 12, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Abbott Alinity m HPV AMP Kit, used with the Alinity m System, product codes: a) REF 09N15-095; b) REF 09N15-090; c) REF 09N15-091
Reason
Abbott has identified an increase of incidences regarding Error Code (EC) 9198 (Positive control is non-reactive) while using the Alinity m HR HPV AMP Kit and Alinity m STI AMP Kit. Certain invalidated positive assay controls can be traced to iron leaching into the Alinity m Lysis Solution from the lysis transfer pump in the Alinity m System.
Classification
Class II
Recalling firm
Abbott Molecular, Inc.
Quantity
13318 units
Agency-reported status
Ongoing
Source record identifier
Z-0233-2025

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilitya) REF 09N15-095, UDI/DI 00884999048614; b) REF 09N15-090, UDI/DI 00884999047921; c) REF 09N15-091, UDI/DI 00884999049529; All Lots

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide distribution: US (nationwide) to states of: AZ, CA, CO, DC, FL, GA, IN, IL, LA, MA, MI, MN, MO, NC, NE, NJ, NY, OR, PA, SC, TN, TX, VA, WA, and WI; and foreign (OUS) countries of: AUSTRIA, BELGUIM, BULGARIA, CAMBODIA, CANADA, COSTA RICA, CROATIA, CZECH REPUBLIC, DOMINICAN REP., EL SALVADOR, ESTONIA, FINLAND, FRANCE, GERMANY, GREECE, IRELAND, ISRAEL, ITALY, KAZAKHSTAN, KENYA, LUXEMBOURG, MALAWI, MALAYSIA, MONTENEGRO, MOROCCO, NETHERLANDS, NORWAY, POLAND, PORTUGAL, ROMANIA, SAUDIA ARABIA, SINGAPORE, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TANZANIA, THAILAND, UNITED KINGDOM, AND VIETNAM.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Nov 6, 2024
Recall initiation date
Sep 12, 2024
Center classification date
Oct 28, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.