ClariTEE Probe- Miniature transesophageal echo ultrasound probe for use with ImaCor ultrasound diagnostic platforms. ... — FDA record Z-0231-2025

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Sep 16, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
ClariTEE Probe- Miniature transesophageal echo ultrasound probe for use with ImaCor ultrasound diagnostic platforms. Model/Catalog Number: CLT-010, CLT-010-1
Reason
Epoxy used to seal the edge of the transducer into its paddle exceeded specification. This variability could attenuate the ultrasound signal with a potential impact on the quality of images obtained from the probes.
Classification
Class II
Recalling firm
ImaCor Inc.
Quantity
73 units
Agency-reported status
Ongoing
Source record identifier
Z-0231-2025

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI - Primary - 00861589000108 Lot numbers: 20250509 and 20250823

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

FL ,GA, NJ, TN

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Nov 6, 2024
Recall initiation date
Sep 16, 2024
Center classification date
Oct 25, 2024