LOW PROF LCKNG SCREW, F/IM NAIL 5.0MM / L, 46MM / XL25/ STER. Product Code: 04.045.346TS. Used in RFN-ADVANCEDTM Retr... — FDA record Z-0228-2026

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Sep 26, 2025. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
LOW PROF LCKNG SCREW, F/IM NAIL 5.0MM / L, 46MM / XL25/ STER. Product Code: 04.045.346TS. Used in RFN-ADVANCEDTM Retrograde Femoral Nailing (RFNA) System and the Tibial Nail Advanced System.
Reason
Although the etched length on the screw is correct, the length marked on the external packaging is incorrect. Specifically, 42mm screws were packaged and labeled as 46mm, while 46mm screws were packaged and labeled as 42mm.
Classification
Class II
Recalling firm
Synthes (USA) Products LLC
Quantity
120 units
Agency-reported status
Ongoing
Source record identifier
Z-0228-2026

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityProduct Code: 04.045.346TS. GTIN: 10886982301413. Lot Number: 31756P7. Expiration Date: 08/31/2025

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide - US Nationwide distribution in the states of AL, AR, AZ, CA, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MN, MO, MS, NJ, NM, NY, OH, TN, TX and the country of Canada.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Oct 29, 2025
Recall initiation date
Sep 26, 2025
Center classification date
Oct 21, 2025

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records