TufTex Embolectomy Catheter, Model No. This catheter has a latex balloon and a single lumen extrusion. It is inserted... — FDA record Z-0228-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Sep 30, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
TufTex Embolectomy Catheter, Model No. This catheter has a latex balloon and a single lumen extrusion. It is inserted using a guidewire or introducer. The devices affected by this recall are 3F and 4F..
Reason
Three lots tied by two trainees were released without performing the required pull test. If tied with inadequate strength, the proximal ligature could slip toward the tip of the catheter and block the inflation hole, leading to the risk of the tip breaking off, and potentially becoming lost in the blood vessel.
Classification
Class II
Recalling firm
LeMaitre Vascular, Inc.
Quantity
72 (OUS only)
Agency-reported status
Ongoing
Source record identifier
Z-0228-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityLots XSL0097, XSL0098, and XSL0099

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

No US distribution. Foreign distribution to United Kingdom, Germany, France, Sweden, Austria, and Italy.

No US distribution state could be established from this record. Check the official notice.

Dates in the source

Report date
Nov 23, 2022
Recall initiation date
Sep 30, 2022
Center classification date
Nov 11, 2022