Elekta Disposable Biopsy Needle Kit for Leksell Stereotactic System, REF 911933, Diameter / 14 Gauge, containing 6 in... — FDA record Z-0227-2025

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Sep 25, 2024. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Elekta Disposable Biopsy Needle Kit for Leksell Stereotactic System, REF 911933, Diameter / 14 Gauge, containing 6 inner boxes labeled REF 911761, Disposable Biopsy Needle Kit for Leksell Stereotactic System. Each inner box of REF 911761 contains a Tyvek bag labeled REF 912465, Disposable Biopsy Needle and Aspirator Tube. The system is used for for localization and diagnosis of intracranial disorders and their surgical treatment including radiosurgery and stereotactic radiation therapy.
Reason
Elekta has become aware that Disposable Biopsy Needles can contain some microscopic debris on the inside of the biopsy needle. The material in the debris is stainless steel, same material as the biopsy needle. No debris has been found on the outer parts of the biopsy needles.
Classification
Class I
Recalling firm
Elekta Instrument AB
Quantity
111 kits (666 needles)
Agency-reported status
Ongoing
Source record identifier
Z-0227-2025

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityBatch Number 873874875R Package including 6 needles: UDI: (01)0 7340048 30034 6 (17) 260301 (10) 873874875R (240) 911933 Package including 1 needle: UDI: (01)0 7340048 30800 7 (17) 260301 (10) 873874875R (240) 911761 Inner sterile bag including 1 needle: UDI: (01)0 7340048 30801 4 (17) 260301 (10) 873874875R (240) 912465

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US: AZ, TX, FL, CA, NY, GA, VA, NJ, MO, PA, WA,

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Nov 13, 2024
Recall initiation date
Sep 25, 2024
Center classification date
Nov 6, 2024