Beaver Xstar Crescent Knife, 2.5mm, 55 degrees, bevel down. Intended for or cataract surgery. REF#: 373809 — FDA record Z-0227-2022

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Oct 8, 2021. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Beaver Xstar Crescent Knife, 2.5mm, 55 degrees, bevel down. Intended for or cataract surgery. REF#: 373809
Reason
Product packaged as Beaver Xstar Crescent Knife, 2.5mm, 55 degrees bevel-down blades contain bevel-up blades
Classification
Class II
Recalling firm
Beaver Visitec International, Inc.
Quantity
5950 eaches
Agency-reported status
Ongoing
Source record identifier
Z-0227-2022

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityLOT#: 6000842, 6002045, 6011609, 6034238, 6040517, 6042907. UDI: " Primary: 00886158000594 " Secondary: 30886158000595 " Tertiary: 50886158000599

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Nationwide Foreign: Australia Austria France Germany Netherlands Spain Japan Russia South Africa UK

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Nov 24, 2021
Recall initiation date
Oct 8, 2021
Center classification date
Nov 16, 2021

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records