Philips Laser System, REF: LAS-100. Used in minimally invasive interventional procedures within the cardiovascular sy... — FDA record Z-0221-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Oct 14, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Philips Laser System, REF: LAS-100. Used in minimally invasive interventional procedures within the cardiovascular system, and for the removal of problematic pacemaker and defibrillator cardiac leads.
Reason
Laser System may detect Error 106 or Error 108 System Failures that will cause the system to enter non-recoverable safe state. The user will not be able to proceed until the error is cleared. May result in initiation of treatment delay and/or treatment beyond initial scope and/or unable to treat patient.
Classification
Class II
Recalling firm
Spectranetics Corporation
Quantity
113
Agency-reported status
Ongoing
Source record identifier
Z-0221-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI-00813132027452, Software Version: 0.5.0.3, Serial Numbers: 100000, 100001, 100002, 100003, 100004, 100005, 100006, 100007, 100008, 100009, 100010, 100011, 100012, 100013, 100014, 100015, 100016, 100017, 100018, 100019, 100020, 100021, 100022, 100023, 100024, 100025, 100026, 100027, 100028, 100029, 100030, 100033, 100034, 100035, 100036, 100037, 100038, 100039, 100040, 100041, 100042, 100043, 100044, 100045, 100046, 100047, 100048, 100049, 100050, 100051, 100052, 100053, 100054, 100055, 100058, 100059, 100060, 100061, 100062, 100063, 100064, 100065, 100066, 100067, 100068, 100069, 100070, 100071, 100072, 100073, 100074, 100075, 100076, 100077, 100078, 100079, 100080, 100081, 100082, 100083, 100084, 100085, 100086, 100087, 100088, 100089, 100090, 100091, 100092, 100093, 100094, 100095, 100096, 100097, 100098, 100099, 100100, 100101, 100102, 100103, 100104, 100105, 100106, 100107, 100108, 100112, 100113, 100114, 100115, 100116, 100117, 100118, 100120

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US: LA, MO, TN, CA, NC, PA, HI, AZ, CO, OK, FL, TX, MN, MI, VA, CT, NY, MD, IL, NJ, MA, KS, ME, AR, IA, IN, GA, ID, OR. OUS: DE, GB, ES, SE, IT, RS, GR, IN, FR, IL, DK, SA

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Nov 23, 2022
Recall initiation date
Oct 14, 2022
Center classification date
Nov 11, 2022

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.