RX Imola-IVD Clinical Chemistry Analyzer Catalog Number: RX4900. RX4900R (refurbished) — FDA record Z-0200-2024
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Sep 6, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- RX Imola-IVD Clinical Chemistry Analyzer Catalog Number: RX4900. RX4900R (refurbished)
- Reason
- RX Parameter rerun settings (Sample Volume Low/High), and Dilutions used for Auto reruns, not meeting sample re-run conditions when samples are above or below assay range and may result i issues with the re-run feature conditions when incorrectly. Potential to report incorrectly.
- Classification
- Class II
- Recalling firm
- Randox Laboratories Ltd.
- Quantity
- 35 units
- Agency-reported status
- Ongoing
- Source record identifier
- Z-0200-2024
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | GTIN: 05055273206104 Lot/Batch Number: (1) 7201-0747 7201-0423 7201-0417 7201-0353 7255-0198 7255-0172 7255-0171 7255-0129 7201-1085 7201-1086 7201-1079 7201-1080 7201-0982 7201-0977 7201-0970 7201-0950 7201-0924 7201-0919 7201-0901 7201-0849 7201-0827 7201-0801 7201-0768 7201-0606 7201-0562 7201-0535 7201-0531 7201-0482 (2)¿¿¿¿¿¿¿¿¿¿¿¿¿ 7201-0776 7201-0367 7201-0343¿¿¿¿¿¿¿¿¿¿¿¿¿ ¿¿¿¿ 7201-0348 ¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿ |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
US Nationwide distribution including Puerto Rico
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
- Alaska
- Alabama
- Arkansas
- Arizona
- California
- Colorado
- Connecticut
- Delaware
- Florida
- Georgia
- Hawaii
- Iowa
- Idaho
- Illinois
- Indiana
- Kansas
- Kentucky
- Louisiana
- Massachusetts
- Maryland
- Maine
- Michigan
- Minnesota
- Missouri
- Mississippi
- Montana
- North Carolina
- North Dakota
- Nebraska
- New Hampshire
- New Jersey
- New Mexico
- Nevada
- New York
- Ohio
- Oklahoma
- Oregon
- Pennsylvania
- Rhode Island
- South Carolina
- South Dakota
- Tennessee
- Texas
- Utah
- Virginia
- Vermont
- Washington
- Wisconsin
- West Virginia
- Wyoming
Dates in the source
- Report date
- Nov 8, 2023
- Recall initiation date
- Sep 6, 2023
- Center classification date
- Oct 31, 2023
FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.