Arietta 65 Ultrasound Software Versions: V1.0 through V4.0.0 — FDA record Z-0192-2022

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Oct 14, 2021. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Arietta 65 Ultrasound Software Versions: V1.0 through V4.0.0
Reason
When the ultrasound system is used in combination with a Fujifilm (formally Hitachi) transducer, there is an error in the calculation formula of Effective Regurgitant Orifice Area (EROA) when you perform MR Flow combined with Left Ventricular Outflow Tract Flow (LVOT Flow) after performing an echocardiogram. The incorrect value is displayed. In addition, the error appears in the instruction manual
Classification
Class II
Recalling firm
FujiFilm Healthcare Americas Corporation
Quantity
48 US
Agency-reported status
Ongoing
Source record identifier
Z-0192-2022

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilitySerial Numbers/UDI: G3024453 (01)04562122959278(11)191029(21)G3024453 G3024456 (01)04562122959278(11)191029(21)G3024456 G3042466 (01)04562122959278(11)191115(21)G3042466 G3042471 (01)04562122959278(11)191115(21)G3042471 G3042477 (01)04562122959278(11)191115(21)G3042477 G3042478 (01)04562122959278(11)191115(21)G3042478 G3042484 (01)04562122959278(11)191115(21)G3042484 G3042486 (01)04562122959278(11)191115(21)G3042486 G3042487 (01)04562122959278(11)191115(21)G3042487 G3042496 (01)04562122959278(11)191115(21)G3042496 G3042499 (01)04562122959278(11)191115(21)G3042499 G3042501 (01)04562122959278(11)191115(21)G3042501 G3042576 (01)04562122959278(11)191220(21)G3042576 G3042578 (01)04562122959278(11)191220(21)G3042578 G3042583 (01)04562122959278(11)191220(21)G3042583 G3042584 (01)04562122959278(11)191205(21)G3042584 G3042585 (01)04562122959278(11)191205(21)G3042585 G3042587 (01)04562122959278(11)191205(21)G3042587 G3042588 (01)04562122959278(11)191220(21)G3042588 G3082956 (01)04562122959278(11)200929(21)G3082956 G3085640 (01)04562122959278(11)200929(21)G3085640 G3085644 (01)04562122959278(11)200929(21)G3085644 G3095193 (01)04562122959278(11)201211(21)G3095193 G3095194 (01)04562122959278(11)201211(21)G3095194 G3095195 (01)04562122959278(11)201211(21)G3095195 G3095197 (01)04562122959278(11)201211(21)G3095197 G3095202 (01)04562122959278(11)201211(21)G3095202 G3095445 (01)04562122959278(11)201228(21)G3095445 G3111596 (01)04562122959278(11)210401(21)G3111596 G3111597 (01)04562122959278(11)210401(21)G3111597 G3111598 (01)04562122959278(11)210401(21)G3111598 G3111605 (01)04562122959278(11)210402(21)G3111605 G3111606 (01)04562122959278(11)210402(21)G3111606 G3114264 (01)04562122959278(11)210428(21)G3114264 G3114267 (01)04562122959278(11)210428(21)G3114267 G3114366 (01)04562122959278(11)210514(21)G3114366 G3114368 (01)04562122959278(11)210514(21)G3114368 G3114372 (01)04562122959278(11)210517(21)G3114372 G3114373 (01)04562122959278(11)210517(21)G3114373 G3114376 (01)04562122959278(11)210517(21)G3114376 G3114379 (01)04562122959278(11)210517(21)G3114379 G3114399 (01)04562122959278(11)210518(21)G3114399 G3114897 (01)04562122959278(11)210519(21)G3114897 G3114902 (01)04562122959278(11)210519(21)G3114902 G3118967 (01)04562122959278(11)210528(21)G3118967 G3118968 (01)04562122959278(11)210528(21)G3118968

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Nov 10, 2021
Recall initiation date
Oct 14, 2021
Center classification date
Nov 2, 2021

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.