Arietta 750 Ultrasound Software Version: V1.0.0 through V2.1.3 — FDA record Z-0189-2022

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Oct 14, 2021. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Arietta 750 Ultrasound Software Version: V1.0.0 through V2.1.3
Reason
Diagnostic ultrasound system with the specified software version and used in conjunction with a Fujifilm (formally Hitachi) transducer does not display the measurement results correctly with two cardiology measurement functions: Mitral Regurgitation Flow (MR Flow) Measurement; Mitral Valve (MV) Measurement. In addition there is an error in the instruction manual.
Classification
Class II
Recalling firm
FujiFilm Healthcare Americas Corporation
Quantity
15 units
Agency-reported status
Ongoing
Source record identifier
Z-0189-2022

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilitySerial Numbers: G3075073 (01)04562122959841(11)201214(21)G3075073; G3110777 (01)04562122959841(11)210421(21)G3110777; G3110771 (01)04562122959841(11)210421(21)G3110771; G3113780 (01)04562122959841(11)210510(21)G3113780; G3113778 (01)04562122959841(11)210510(21)G3113778; G3113779 (01)04562122959841(11)210510(21)G3113779; G3113777 (01)04562122959841(11)210510(21)G3113777; G3071857 (01)04562122959841(11)210420(21)G3071857; G3108384 (01)04562122959841(11)210402(21)G3108384; G3114220 (01)04562122959841(11)210510(21)G3114220; G3110776 (01)04562122959841(11)210421(21)G3110776; G3108385 (01)04562122959841(11)210402(21)G3108385; G3114223 (01)04562122959841(11)210518(21)G3114223; G3108401 (01)04562122959841(11)210420(21)G3108401; G3114959 (01)04562122959841(11)210524(21)G3114959

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Nationwide

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Nov 10, 2021
Recall initiation date
Oct 14, 2021
Center classification date
Nov 2, 2021