Brand Name: MEGA Ballast Distal Access Platform Product Name: Mega Ballast Model/Catalog Number: MEGABT09180, MEGABT0... — FDA record Z-0188-2026

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Sep 25, 2025. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Brand Name: MEGA Ballast Distal Access Platform Product Name: Mega Ballast Model/Catalog Number: MEGABT09180, MEGABT09190, MEGABT091100, MEGABT091110 Software Version: N/A Product Description: The Mega Ballast 091 is designed with a semi-rigid proximal shaft and a highly flexible distal shaft. The design facilitates the advancement of the Mega Ballast to access the tortuous vasculature over a guidewire and a respective dilator. The outer surface of the Mega Ballast is coated with a hydrophilic coating to facilitate lubricity from tip to 40 cm proximal. The proximal end of the Mega Ballast incorporates a Luer fitting hub for the attachment of accessories. The Mega Ballast is provided sterile, non-pyrogenic, and is intended for single use only. Component: Not a component
Reason
Due to an identified manufacturing defect involving coating material lifting within the catheter's distal inner diameter/lumen.
Classification
Class I
Recalling firm
BALT USA, LLC
Quantity
207 catheters
Agency-reported status
Ongoing
Source record identifier
Z-0188-2026

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityModel Numbers and UDI-DI codes: Model No MEGABT09180 (00840303711780) Model No MEGABT09190 (00840303711797) Model No MEGABT091100 (00840303711803) Model No MEGABT091110 (00840303711810) Lot Numbers: F250600687, F250600686, F250600685, F250600455, F250600346, F250600345, F250600010, F250501434, F250501346, F250501091, F250500824, F250400219, F250400024, F250301020, F250300297, F250100956

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US distribution to states of: AL, CA, CO, CT, FL, IL, KY, MA, NY, TN, TX, and UT.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Nov 5, 2025
Recall initiation date
Sep 25, 2025
Center classification date
Oct 24, 2025

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.