Medical convenience kits UDPG88W PICC G-TUBE PACK-230209 — FDA record Z-0186-2026

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Sep 4, 2025. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Medical convenience kits UDPG88W PICC G-TUBE PACK-230209
Reason
Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.
Classification
Class II
Recalling firm
American Contract Systems Inc.
Quantity
32,433 total
Agency-reported status
Ongoing
Source record identifier
Z-0186-2026

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDPG88W UDI-DI 191072216359 lot 7935611 bag serial number N/A * If bag serial number is listed as "N/A", then whole lot is affected

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution in the states of SD, IA, MN, WA, IL.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Oct 22, 2025
Recall initiation date
Sep 4, 2025
Center classification date
Oct 16, 2025

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.