Recall archive / FDA device enforcement
Medical convenience kits Item Number/Description BUCY78E CYSTO PACK CMCP10X CYSTO PACK - 205972 IHCY10Y CYSTO PACK MH... — FDA record Z-0174-2026
Reported ongoing
Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.
Source date Sep 4, 2025. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.
Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.
What the record says
- Product
- Medical convenience kits Item Number/Description BUCY78E CYSTO PACK CMCP10X CYSTO PACK - 205972 IHCY10Y CYSTO PACK MHCY35AE CYSTO PACK- 212487 PLBY52E CYSTO PACK UDCY31Z CYSTO PACK - 206047 UICY28L MOR CYSTO PACK
- Reason
- Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.
- Classification
- Class II
- Recalling firm
- American Contract Systems Inc.
- Quantity
- 32,433 total
- Agency-reported status
- Ongoing
- Source record identifier
- Z-0174-2026
What the recall classes mean. Classification is an agency assessment, not individual medical advice.
Lot and product identifiers
| Label field | Details supplied by the agency |
|---|---|
| Lot, code or applicability | BUCY78E UDI-DI 191072235350 Lot 8531411 Bag serial number 17705421; CMCP10X UDI-DI 191072228215 Lot 8392411 Bag serial number 17622370; IHCY10Y UDI-DI 191072159922 Lot 8026111 Bag serial number 17499185 Lot 8346211 Bag serial number 17692649 Lot 8444211 Bag serial number 17681124; MHCY35AE UDI-DI 191072228864 Lot 7990511 Bag serial number N/A; PLBY52E UDI-DI 191072141415 Lot 8016811 Bag serial number 17435184; UDCY31Z UDI-DI 00191072213990 Lot 8043211 Bag serial number 17483121 17483122 17483143 17483144 17485283; UICY28L UDI-DI 191072205537 Lot 8217311 Bag serial number 17598793 * If Bag serial number is listed as "N/A", then whole lot is affected |
Match all identifying details in the current notice. A brand name alone is not enough.
Distribution
US Nationwide distribution in the states of SD, IA, MN, WA, IL.
States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.
- Alaska
- Alabama
- Arkansas
- Arizona
- California
- Colorado
- Connecticut
- Delaware
- Florida
- Georgia
- Hawaii
- Iowa
- Idaho
- Illinois
- Indiana
- Kansas
- Kentucky
- Louisiana
- Massachusetts
- Maryland
- Maine
- Michigan
- Minnesota
- Missouri
- Mississippi
- Montana
- North Carolina
- North Dakota
- Nebraska
- New Hampshire
- New Jersey
- New Mexico
- Nevada
- New York
- Ohio
- Oklahoma
- Oregon
- Pennsylvania
- Rhode Island
- South Carolina
- South Dakota
- Tennessee
- Texas
- Utah
- Virginia
- Vermont
- Washington
- Wisconsin
- West Virginia
- Wyoming
Dates in the source
- Report date
- Oct 22, 2025
- Recall initiation date
- Sep 4, 2025
- Center classification date
- Oct 16, 2025
FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.