BD Trucount Tubes (Cat. No. 340334), used for determining absolute counts of leucocytes in blood. BD Trucount tubes a... — FDA record Z-0171-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Oct 3, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
BD Trucount Tubes (Cat. No. 340334), used for determining absolute counts of leucocytes in blood. BD Trucount tubes are designed for use with in vitro diagnostic products such as BD Tritest reagents, and a suitably equipped flow cytometer. BD Trucount tubes can be used with the BD FACS Loader.
Reason
Due to partial label detachment on tubes. This can cause tubes to become stuck together in the pouch or cause automation errors when used with the automated loader for the wash assistant. Firm root cause investigation confirmed that label detachment is due to an adhesive change by a tier 2 supplier.
Classification
Class II
Recalling firm
Becton, Dickinson and Company, BD Biosciences
Quantity
52,570 tubes (corrected on 11/04/2022)
Agency-reported status
Ongoing
Source record identifier
Z-0171-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityAdded/Corrected as of 11/04/2022: Catalog Number: 340334 UDI-DI Code: 00382903403349 Lot Numbers: 2031144 2034642 2035110 2035121 2060660 2073297 2073298 2095186 2101458 2108676 2112862 2152055 2152064 2152068 2152075 2152080 2203024 2208091 2215812 2235269 2223229 2236782 2236934 2236936 2236937 2236783 2237006 2237008 2237012 2237014 2237015 2237016

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

U.S.: AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, NC, ND, NE, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SD, TN, TX, UT, VA, VT, WA, and WI O.U.S.: Argentina, Australia, Belgium, Bolivia, Brazil, Canada, Chile, China, Columbia, Costa Rica, Dominican Republic, Ecuador, El Salvador, Guatemala, Hong Kong, India, Japan, Korea, Mexico, Malaysia, New Zealand, Panama, Peru, Philippines, Paraguay, Singapore, Thailand, Taiwan, and Uruguay

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Nov 9, 2022
Recall initiation date
Oct 3, 2022
Center classification date
Nov 3, 2022

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.