Senhance Surgical System, which includes: Manipulator Arm, ISU Configuration ( Catalog Number X9007708), Intelligent ... — FDA record Z-0170-2024

Reported terminated

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Sep 15, 2023. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Senhance Surgical System, which includes: Manipulator Arm, ISU Configuration ( Catalog Number X9007708), Intelligent Surgical Unit (Catalog Number X9007696), and Cockpit, ISU Configuration (Catalog Number X9007707), Software (SW) version 2.7.4; robotic surgical system
Reason
Asensus Surgical has identified a malfunction that has occurred and may recur on the Senhance Surgical System. This issue presented itself as uncontrolled arm motion of the Laparoscope Instrument Actuator (LIA) where the LIA rotated continuously in one direction after the Surgeon removed the engagement of teleoperation on the Senhance System. There was no other uncontrolled arm motion observed.
Classification
Class I
Recalling firm
Asensus Surgical Inc
Quantity
21 units
Agency-reported status
Terminated
Source record identifier
Z-0170-2024

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI/DI 00815440022075, Serial Numbers: SN 00012, SN 00016, SN 00007, SN 00017, SN 00035, SN 00018, SN 00042, SN 00008, SN 00031, SN 00025, SN 00011, SN 00028, SN 00038, SN 00029, SN 00036, SN 00025, SN 00024, SN 00006, SN 00019, SN 00040, SN 00041, loaded with Software (SW) version 2.7.4

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US: NJ, LA, FL, MN, IL, and Germany, Japan, Lithuania

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Nov 8, 2023
Recall initiation date
Sep 15, 2023
Center classification date
Oct 30, 2023
Termination date
Sep 26, 2024