Brand Name: ARVIS Product Name: ARVIS" Shoulder Model/Catalog Number: IN-27300, Arvis Surgical Planning Software Soft... — FDA record Z-0145-2026

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Sep 19, 2025. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Brand Name: ARVIS Product Name: ARVIS" Shoulder Model/Catalog Number: IN-27300, Arvis Surgical Planning Software Software Version: V2025.1.2 Product Description: Intended Use: The ARVIS" system is a computer-controlled navigation system intended to provide intra-operative measurements to the surgeon to aid in selection and positioning of orthopedic implant components. Description: ARVIS Shoulder is a computer-controlled surgical navigation system for shoulder arthroplasty. It aids the surgeon in making intra-operative measurements and locating anatomical structures of the shoulder joint based on the patient s preoperative imaging to assist with selection and positioning of orthopedic implant components. The system consists of software, electronic hardware and surgical instruments. The device s workflow involves CT based preoperative planning followed by intraoperative navigation and execution. The preoperative planning software enables 3D virtual implant positioning based on the patient s CT reconstructed digital bone model and bony landmarks. The shoulder navigation application software then matches the patient s digital bone model and landmarks to the intraoperative landmarks registered by the surgeon, allowing an image-based navigation to follow. The surgeon uses the plan data as guidance to navigate and help position shoulder instruments and implants.
Reason
Complaint identified issue with AI surgical planning software that may result in implant malalignment and/or decrease range of motion.
Classification
Class II
Recalling firm
Kico Knee Innovation Company
Quantity
21 systems
Agency-reported status
Ongoing
Source record identifier
Z-0145-2026

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilitySoftware Version V2025.1.2

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US Nationwide distribution in the states of IN, MN, MO, OH, FL, IL, RI, KY, CA, WI, MI, PA, WA, CO, AL.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Oct 22, 2025
Recall initiation date
Sep 19, 2025
Center classification date
Oct 14, 2025

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.

Related records