Hillrom Hospital Bed Accessory, Watch Care Incontinence Management System for VersaCare Bed Rev. A-J, Product Code/Pa... — FDA record Z-0145-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Sep 30, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
Hillrom Hospital Bed Accessory, Watch Care Incontinence Management System for VersaCare Bed Rev. A-J, Product Code/Part Number P00697901
Reason
Radio frequency emission from a functioning WatchCare device may potentially impact other devices (including, but not limited to telemetry devices, bladder scanner, fetal monitor/doppler, infusion pumps, and insulin pumps).
Classification
Class I
Recalling firm
Baxter Healthcare Corporation
Quantity
129 units
Agency-reported status
Ongoing
Source record identifier
Z-0145-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI/DI 00887761998865, all serial numbers

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

US

No US distribution state could be established from this record. Check the official notice.

Dates in the source

Report date
Nov 9, 2022
Recall initiation date
Sep 30, 2022
Center classification date
Oct 31, 2022