CorVocet Biopsy System, Catalog Code: CORA1410/B, CORA1410S/B, CORA1415/B, CORA1615S/B, CORA2010S/B, CORCA1410S/B, CO... — FDA record Z-0138-2023

Reported ongoing

Status as reported in the agency record. FDA warns that enforcement status may not track the current recall lifecycle; check the current official notice.

Source date Aug 3, 2022. Retrieved Oct 1, 2026. Last updated Oct 1, 2026.

Archive changed Oct 1, 2026. USDA FSIS records are unavailable; check FSIS directly. Source coverage and dates.

What the record says

Product
CorVocet Biopsy System, Catalog Code: CORA1410/B, CORA1410S/B, CORA1415/B, CORA1615S/B, CORA2010S/B, CORCA1410S/B, CORCA1410S/CNB, CORCA1415S/B, CORCA1610/CNB, CORCA1615S/CNB, CORCA1810SB/CNB, CORCA1815SB/CNB, CORCA1820SB/CNB, CORCA2015SB/CNB, CORCA2020SB/CNB
Reason
Biopsy Systems have components that are misaligned due to an issue with the manufacturing assembly equipment, which may lead to the needle ejecting when triggered outside the body, which may result in user or patient needle stick injury.
Classification
Class II
Recalling firm
Merit Medical Systems, Inc.
Quantity
2894
Agency-reported status
Ongoing
Source record identifier
Z-0138-2023

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityCatalog Code-Lot (OUS Distribution Only): CORA1410/B- I2206909, I2318475, I2291980, I2286704; CORA1410S/B- I2206888; CORA1415/B- I2213361; CORA1615S/B- I2114518, H1517631, H1715532, H1756745, H1767115, H1771762, H1776067, H1784861, H1800979, H1808121, H1825693, I1781878, I1818817, I1837234, I1843704, I1855418, I1866359, I1952354, I1958525, I1971876, I1990075, I1998862, I2043630, I2070249, I2075810; CORA2010S/B- I2010414; CORCA1410S/B- I2259093; CORCA1410S/CNB- I2305358; CORCA1415S/B- I2210950; CORCA1610/CNB- I2291983; CORCA1615S/CNB- I2330683; CORCA1810SB/CNB- I2305359; CORCA1815SB/CNB- I2170254, I2205164; CORCA1820SB/CNB- I2325912; CORCA2015SB/CNB- I2318903, I2325926; CORCA2020SB/CNB- I2320880

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide distribution - US National distribution including in the states of FL, IL, OH, PA, CA, WI, VA, AZ, AR, MS, TN, MO, MA, MI, NY, NE, TX, WA, MN, LA, DE, IN, AL, NJ, OR, MD, GA, CT, OK, UT, KS, NC, CO, KY, WV, AK, NV, RI, SC, ND, NH, MT and the countries of Finland, Denmark, United Arab Emirates, Spain, Sweden, Norway, Jordan, Australia, Germany, Austria, Italy, United Kingdom, France, Canada, Mexico, Puerto Rico, Chile, Reunion, Belgium, Switzerland, Japan, South Africa, Qatar, Russian Federation, Hungary, Thailand, Namibia, Colombia, New Zealand, Netherlands, Saudi Arabia, Costa Rica, Oman, Cyprus, Slovenia, Croatia, China.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Nov 2, 2022
Recall initiation date
Aug 3, 2022
Center classification date
Oct 27, 2022