What the record says

Product
Soltive Pro SuperPulsed Laser Systems-Intended for incision, excision, resection, ablation, coagulation, hemostasis, and vaporization of soft tissue Model/Catalog Number: TFL-SLS
Reason
The term Bladder Stone was incorrectly translated in both Spanish and Portuguese to Kidney Stone(Clculo renal) on the systems Graphical User Interface (GUI), a potential to deliver an incorrect or unintended amount of energy to the patients anatomy.
Classification
Class II
Recalling firm
Olympus Corporation of the Americas
Quantity
159 units
Agency-reported status
Ongoing
Source record identifier
Z-0135-2025

Open the official record

What the recall classes mean. Classification is an agency assessment, not individual medical advice.

Lot and product identifiers

Label fieldDetails supplied by the agency
Lot, code or applicabilityUDI 00821925044135 All Serial Numbers

Match all identifying details in the current notice. A brand name alone is not enough.

Distribution

Worldwide distribution - US Nationwide and the countries of AU, CA, CL, DE, HK, JP, SG.

States below are derived from the distribution text, including nationwide wording. They are not inferred from the firm's address.

Dates in the source

Report date
Oct 30, 2024
Recall initiation date
Aug 13, 2024
Center classification date
Oct 18, 2024

FDA company pages use the listing date; enforcement records use initiation dates; CPSC uses recall date; NHTSA uses report-received date. These dates are not interchangeable.